Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27051 |
| Brand Name: | Alaris Plus Volumetric Pumps_BD Swiss |
| Model: | 9002TIG03 - Alaris™ GP Volumetric Pump with Plus software, 9002TIG03 - GAlaris™ GP Guardrails™ Volumetric Pump with Plus software |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Becton Dickinson Ltd Unit 4 / 110 Mays Road Onehunga Auckland 1061 |
| Contact Information: | 0800 572 468 |
| Manufacturer: | BD Switzerland Sarl |
| Issue: | The manufacturer has indentified that over time, the devices' case polymer is sensitive to chemicals used during the cleaning process at point of use. The manufacturer has given guidance on recommended cleaning agents following chemical compatibility studies completed internally. As a result of an internal documentation review, it was identified that Tuffie 5 was included as a Recommended Cleaner in the Directions for Use(DFU)in error, in units distributed betweenJuly 2018 and August 2020. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 15/12/2020 |