Type of Product:Device
Medsafe Reference:27051
Brand Name:Alaris Plus Volumetric Pumps_BD Swiss
Model:9002TIG03 - Alaris™ GP Volumetric Pump with Plus software, 9002TIG03 - GAlaris™ GP Guardrails™ Volumetric Pump with Plus software
Affected:
Software version:
Recalling Organisation:Becton Dickinson Ltd
Unit 4 / 110 Mays Road
Onehunga
Auckland 1061
Contact Information:0800 572 468
Manufacturer:BD Switzerland Sarl
Issue:The manufacturer has indentified that over time, the devices' case polymer is sensitive to chemicals used during the cleaning process at point of use. The manufacturer has given guidance on recommended cleaning agents following chemical compatibility studies completed internally. As a result of an internal documentation review, it was identified that Tuffie 5 was included as a Recommended Cleaner in the Directions for Use(DFU)in error, in units distributed betweenJuly 2018 and August 2020.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/12/2020