Detail of Recall

Type of Product:Device
Medsafe Reference:27033
Brand Name:Nuclear Medicine System, Gamma Camera, Stationary
Model:
Affected:Rotor Bearing Screws for Nuclear Medicine Systems
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems Functional Imaging Haifa
Issue:The manufacturer has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector. As a result of multiple other design redundancies the detector remained secured. As a result of these design redundancies, it is highly improbable that the detector would become unsecure if the bearing screws loosen. In the unlikely event that the bearing screws loosen, and the multiple other design redundancies also fail, the detector could fall during use and lead to potentially life-threatening bodily harm.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/12/2020