Detail of Recall

Type of Product:Device
Medsafe Reference:27021
Brand Name:Hamilton_Humidifier
Model:
Affected:software version HPC 1.10c
Software version:
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:Hamilton Medical AG
Issue:The manufacturer has identified an issue relating to a delayed low water level alarm when starting the humidifier device with an empty chamber. This alarm delay occurs only in the HAMILTON- H900 software version HPC 1.10c. This issue pertaining to a delayed low water level alarm when starting the humidifier device with an empty chamber, has been corrected in the HAMILTON-H900 software version 1.10d.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/12/2020