| Type of Product: | Device |
| Medsafe Reference: | 27021 |
| Brand Name: | Hamilton_Humidifier |
| Model: | |
| Affected: | software version HPC 1.10c |
| Software version: | |
| Recalling Organisation: | InterMed Medical Ltd 71 Apollo Drive Albany AUCKLAND 1311 |
| Contact Information: | 0800 333 444 |
| Manufacturer: | Hamilton Medical AG |
| Issue: | The manufacturer has identified an issue relating to a delayed low water level alarm when starting the humidifier device with an empty chamber. This alarm delay occurs only in the HAMILTON- H900 software version HPC 1.10c. This issue pertaining to a delayed low water level alarm when starting the humidifier device with an empty chamber, has been corrected in the HAMILTON-H900 software version 1.10d. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/12/2020 |