Detail of Recall

Type of Product:Device
Medsafe Reference:27005
Brand Name: IOLMaster
Model:ZEISS IOLMaster 700
Affected:
Software version:
Recalling Organisation:Carl Zeiss (NZ) Ltd
22 Fremlin Place
Avondale
Auckland 1026
Contact Information:09 838 5626
Manufacturer:Carl Zeiss Meditec AG
Issue:The manufacturer has indentified that the ZEISS IOLMaster 700 with software version 1.90.2.09 and 1.90.8.06, when used in conjunction with modality worklist (MWL) functionality, is not per-forming as expected. In rare cases this leads to a false display of patients in the patient manager screen, which can subsequently lead to the implantation of an incorrect intraocular lens (IOL).
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/12/2020