Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:26992
Brand Name:DIB - Carelink - Minimed
Model:CareLink™ Personal v14.1b (MMT-7333) using CareLink™ uploader (ACC-7350), CareLink™ system v3.1b (MMT-7350) using CareLink™ uploader (ACC-7350)
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic MiniMed
Issue:The manufacturer released an update to the CareLink™ uploader on Thursday August 27, 2020. The installation of the new uploader, version 3.1.2.000 or 3.1.5.000, to a computer that was subsequently used to generate a CareLink™ report may result in one of the following scenarios:
1. Some reports may be missing all programmed basal rates in the device. This may result in a blank value on the Device Settings report of the 24-hour basal total.
2. Some reports may only display certain programmed basal rates and some basal rates in the device may be missing from the reports. This may result in a miscalculation on the Device Settings report of the 24-hour basal total.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Healthcare Professional
Date Commenced:8/12/2020