| Type of Product: | Device |
| Medsafe Reference: | 26992 |
| Brand Name: | DIB - Carelink - Minimed |
| Model: | CareLink™ Personal v14.1b (MMT-7333) using CareLink™ uploader (ACC-7350), CareLink™ system v3.1b (MMT-7350) using CareLink™ uploader (ACC-7350) |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic MiniMed |
| Issue: | The manufacturer released an update to the CareLink™ uploader on Thursday August 27, 2020. The installation of the new uploader, version 3.1.2.000 or 3.1.5.000, to a computer that was subsequently used to generate a CareLink™ report may result in one of the following scenarios: 1. Some reports may be missing all programmed basal rates in the device. This may result in a blank value on the Device Settings report of the 24-hour basal total. 2. Some reports may only display certain programmed basal rates and some basal rates in the device may be missing from the reports. This may result in a miscalculation on the Device Settings report of the 24-hour basal total. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Healthcare Professional |
| Date Commenced: | 8/12/2020 |