Detail of Recall

Type of Product:Device
Medsafe Reference:26973
Brand Name:cobas
Model:cobas® 8000 Core Unit, cobas pro sample supply unit, cobas c 513 analyse, cobas® 8000 modular analyser series.
Affected:
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:The manufacturer received a customer complaint for the cobas® 8000 modular analyser series where due to a software limitation; the system stopped reading barcoded samples with loss of Utility settings. Two similar complaints were received for cobas pro integrated solutions. While no impact on patient results occurred, internal investigations revealed that for some settings under specific conditions, the occurrence of this situation may remain undetected and leads to deactivation of clotting and foam data flags.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/3/2023