| Type of Product: | Device |
| Medsafe Reference: | 26958 |
| Brand Name: | V680 Ventilator |
| Model: | All V680 ventilators |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Respironics California LLC |
| Issue: | During dual-limb invasive ventilation, a patient cough of sufficient magnitude to drive circuit pressures above 95 cmH2O for longer than 150 milliseconds may cause a Vent Inoperative 1008: Machine and Proximal Pressure Sensors Failed alarm, which will cause the V680 ventilator to cease therapy. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/12/2020 |