Detail of Recall

Type of Product:Device
Medsafe Reference:26958
Brand Name:V680 Ventilator
Model:All V680 ventilators
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Respironics California LLC
Issue:During dual-limb invasive ventilation, a patient cough of sufficient magnitude to drive circuit pressures above 95 cmH2O for longer than 150 milliseconds may cause a Vent Inoperative 1008: Machine and Proximal Pressure Sensors Failed alarm, which will cause the V680 ventilator to cease therapy.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/12/2020