Detail of Recall

Type of Product:Device
Medsafe Reference:26938
Brand Name:ADVIA Chemistry System
Model:
Affected:Gamma-Glutamyl Transferase - All lots, Gamma-Glutamyl Transferase - All lots
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue: The manufacturer has confirmed that the ADVIA Chemistry and Atellica CH Gamma-Glutamyl Transferase (GGT) reagent may demonstrate a coefficient of variation (%CV) outside of the performance data for repeatability (within-run) listed in the ADVIA Chemistry and Atellica CH GGT Instructions For Use (IFU) on some analyzers. The increased imprecision is only observed at low concentrations of approximately 27-42 U/L. The imprecision observed at these low concentrations does not impact the mean recovery of quality control material.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/11/2020