Detail of Recall

Type of Product:Device
Medsafe Reference:26932
Brand Name:cobas z480
Model:
Affected:a dirty lens.
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:The manufacturer has advised that the availability of the Maintenance Checker software for the cobas z 480 analyzer, which must be installed on all cobas z 480 analyzers running any cobas® 4800 assay in the IVD workflow. This service tool aids in identifying conditions that could contribute to the generation of false positive results due dirty optical lenses and spatial cross-talk. Stringent criteria have been established to identify a dirty lens before false positive results are generated.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/2/2022