Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:26900
Brand Name:TDM assays
Model:
Affected:TDM assays Vancomycin, Phenytoin and Phenobarbital
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:Internal experiments confirmed that for the TDM assays Vancomycin, Phenytoin and Phenobarbital after a time period without pipetting from the c pack green (reagent cassette) one of the first two measurements is potentially affected by low recovery.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/11/2020