| Type of Product: | Device |
| Medsafe Reference: | 26900 |
| Brand Name: | TDM assays |
| Model: | |
| Affected: | TDM assays Vancomycin, Phenytoin and Phenobarbital |
| Software version: | |
| Recalling Organisation: | Roche Diagnostics (NZ) Ltd 15 Rakino Way Mt Wellington AUCKLAND 1644 |
| Contact Information: | 0800 363 622 |
| Manufacturer: | Roche Diagnostics GmbH |
| Issue: | Internal experiments confirmed that for the TDM assays Vancomycin, Phenytoin and Phenobarbital after a time period without pipetting from the c pack green (reagent cassette) one of the first two measurements is potentially affected by low recovery. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/11/2020 |