Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 26889 |
| Brand Name: | General Ultrasound Imaging System |
| Model: | |
| Affected: | Software versions affected worldwide: 1.0.x, 1.1.x, 1.2.x, 1.3.x, 1.4.x, 1.5.x, 1.7.x, 1.8.x, 2.0.x, 3.0.x, 4.0.x, 5.0, 5.0.1 |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Ultrasound LLC |
| Issue: | The manufacturer has determined that with certain workflows there is potential for incorrect patient data to be displayed and saved into an exam. An error may occur if you attempt to correct the mistake by editing the patient information or selecting a different patient from the worklist. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/11/2020 |