Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:26884
Brand Name:Mobile Digital X Ray System
Model:
Affected:MobileDiagnost wDR Rel 2.2
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:The manufacturer has identified that if a single fault condition such as earth/ground connection failure were to occur while the MobileDiagnost wDR Rel 2.2 is taking an exposure while connected to mains power, the leakage current (touch current) will exceed the 500 µA permitted under standard AS/NZ 3551 2012 Appendix B- Electrical Safety testing (SECTION B10).
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/11/2020