Detail of Recall

Type of Product:Device
Medsafe Reference:26880
Brand Name:HLS Set
Model:BE-HLS 7050 Article no 70104.7753, BE-HLS 5050 Article no70104.8127
Affected:
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet Cardio Pulmonary GmbH
Issue:During verification testing and a review of customer complaints for the HLS Set Maquet Cardiopulmonary (MCP) identified a potential impairment of the sterile packaging barrier. Under the transport simulation verification tests, movements of the device and its accessories inside the plastic tray have been identified.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/11/2020