Detail of Recall

Type of Product:Device
Medsafe Reference:26851
Brand Name:Artis Q, Q.zen, zeego, zee systems
Model:Artis zee/ Q/ Q.zen systems with software version VD11C, VD11D and VD11E
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:If fluoroscopy exam set parameters like frame rate, pulse rate or zoom parameters are changed during a procedure while keeping the fluoroscopy pedal of the footswitch pressed, the patient entrance dose rate (Air Kerma Rate, AKR) may exceed regulatory limits.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/11/2020