Detail of Recall

Type of Product:Device
Medsafe Reference:26840
Brand Name:ARCHITECT EBV VCA IgM Calibrator
Model:
Affected:Lot number - 08159BE00
Software version:
Recalling Organisation:Abbott Diagnostics Division
Ground Floor, Building D
4 Pacific Rise
Mt Wellington
AUCKLAND 1640
Contact Information:0800 656 233
Manufacturer:Abbott GmbH & Co. KG
Issue:The manufacturer has identified the ARCHITECT EBV VCA IgM Calibrator lot 08159BE00 may show reduced RLU (Relative Light Units) signal. This reduced RLU signal may cause a decrease in cutoff RLU values in assay calibration resulting in:
Upward shift in Quality Control S/CO values
Non-reactive to false grayzone patient results in the range of 0.50 - 0.56 S/CO
Grayzone to false reactive patient results in the range of 1.00 - 1.13 S/CO
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory