Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:26829
Brand Name:Memometal - Fixation device, internal, staple
Model:
Affected:Multiple lots
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Memometal Technologies SAS
Issue:It has been determined that paediatric patients with a body weight of less than 20 kg should have a maximum of two (2) EasyClip staples implanted to remain within the acceptable Margin of Safety. Therefore, the current IFU will be updated
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Healthcare Professional
Date Commenced:3/11/2020