Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 26829 |
| Brand Name: | Memometal - Fixation device, internal, staple |
| Model: | |
| Affected: | Multiple lots |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Memometal Technologies SAS |
| Issue: | It has been determined that paediatric patients with a body weight of less than 20 kg should have a maximum of two (2) EasyClip staples implanted to remain within the acceptable Margin of Safety. Therefore, the current IFU will be updated |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Healthcare Professional |
| Date Commenced: | 3/11/2020 |