Detail of Recall

Type of Product:Device
Medsafe Reference:26784
Brand Name:Anaesthesia System
Model:for Flow-c, Flow-e and Flow-i
Affected:
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet Critical Care AB
Issue:In a few occasions, a pilot valve has been found to malfunction, as the pilot valve didn't open when supposed to. This is related to bad soldering in the concerned component found in certain identified batches from the sub-supplier.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/10/2020