Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 26784 |
| Brand Name: | Anaesthesia System |
| Model: | for Flow-c, Flow-e and Flow-i |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Getinge Australia Pty Ltd (NZ) Unit D, Ground Floor, Building A, Millenium Centre 600 Great South Rd Ellerslie Auckland 1051 |
| Contact Information: | 09 272 9039 |
| Manufacturer: | Maquet Critical Care AB |
| Issue: | In a few occasions, a pilot valve has been found to malfunction, as the pilot valve didn't open when supposed to. This is related to bad soldering in the concerned component found in certain identified batches from the sub-supplier. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 28/10/2020 |