Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:26756
Brand Name:CS - CoreValve Evolut R_Low Risk
Model:EVOLUTR-23, EVOLUTR-26, EVOLUTR-29, EVOLUTR-34
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Core Valve LLC
Brand Name:CS - CoreValve Evolut PRO_Low Risk
Model:EVOLUTPRO-23, EVOLUTPRO-26, EVOLUTPRO-29
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Core Valve LLC
Brand Name:SH - Medtronic Evolut™ PRO+ System
Model:EVPROPLUS-23US, EVPROPLUS-26US, EVPROPLUS-29US, EVPROPLUS-34US
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Core Valve LLC
Issue:The manufacturer has received reports of Evolut™ valve leaflet damage occurring following PID at a rate of 0.020%*. These complaints of damage to the bioprosthetic leaflets resulted in moderate or severe aortic insufficiency which were detected acutely or during follow up. These reported events required re-intervention (77%), conversion to surgery (19%), re-intervention followed by surgery (2%), or were treated conservatively (2%).
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Healthcare Professional
Date Commenced:23/10/2020
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /