Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:26727
Brand Name:DXU Bladderscanner Ultrasound Probes
Model:BLADDERSCAN PRIME PLUS PROBE HANDLE
Affected:
Software version:
Recalling Organisation:Paragon Care Group New Zealand Limited
69 Elizabeth Knox Place
Glen Innes
AUCKLAND
Contact Information:0508 654 258
Manufacturer:Verathon Inc
Issue:The manufacturer has identified an issue with the handle of the Prime Plus probe. On May 27, 2020, a Verathon supplier initiated an unauthorized manufacturing change. This manufacturing change caused the Prime Plus probe handle plastic to be exposed to excessive heat during the molding process. This created the potential for the handle to become brittle and crack
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory