Detail of Recall

Type of Product:Device
Medsafe Reference:26712
Brand Name:Geistlich Bio-Oss Pen
Model:
Affected:Batches as Specified* �] Sizes 0.5cc, 1cc, and 1.5cc
Software version:
Recalling Organisation:Geistlich Pharma Zealand Limited
8A Rennie Drive
Airport Oaks
Auckland
Contact Information:0800 500 043
Manufacturer:Geistlich Pharma AG
Issue:The manufacturer has received reports that it was difficult to expel the Geistlich Bio� Oss granules from the pen. It was also reported that the resistance could suddenly drop and the granules may be expelled all at once. In one of these reported cases ,when the device was used for sinus floor elevation, a rupture of the Schneiderian membrane occurred.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/10/2020