| Type of Product: | Device |
| Medsafe Reference: | 26712 |
| Brand Name: | Geistlich Bio-Oss Pen |
| Model: | |
| Affected: | Batches as Specified* �] Sizes 0.5cc, 1cc, and 1.5cc |
| Software version: | |
| Recalling Organisation: | Geistlich Pharma Zealand Limited 8A Rennie Drive Airport Oaks Auckland |
| Contact Information: | 0800 500 043 |
| Manufacturer: | Geistlich Pharma AG |
| Issue: | The manufacturer has received reports that it was difficult to expel the Geistlich Bio� Oss granules from the pen. It was also reported that the resistance could suddenly drop and the granules may be expelled all at once. In one of these reported cases ,when the device was used for sinus floor elevation, a rupture of the Schneiderian membrane occurred. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |