Detail of Recall

Type of Product:Device
Medsafe Reference:26704
Brand Name:ABL800 analyzers with FLEXQ module
Model:All Serial Numbers
Affected:
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:The manufacturer has identified an error that may lead to potentially assigning erroneous measurement results belonging to the current patient to another patient. In a worst-case scenario, where a critical parameter, such as K+ or pO2 is interrogated, based upon wrongful clinical diagnosis inferred from the mixed-up results, a critically ill patient may be subjected to lack of vital treatment.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/10/2020