Detail of Recall

Type of Product:Device
Medsafe Reference:26702
Brand Name:Heater Cooler Unit
Model:
Affected:All HCU 40 units below S/N: 90441922
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet Cardio Pulmonary GmbH
Issue:The manufacturer has received isolated complaints in relation to heat damage of the cable insulation at the crimp connections of the power cable and the internal electrical connection (L and N) cables between the main circuit breaker and line filter of the HCU 40 Heater-Cooler Unit.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/10/2020