Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 26692 |
| Brand Name: | RaySearch_Radiation therapy treatment planning system |
| Model: | |
| Affected: | RayStation 4, 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, 10A, RayPlan 1, 2, 7, 8A, 8B, 9A, 9B, 10A, including all service packs |
| Software version: | |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | The manufacturer has identified an issue found with the DICOM import in RayStation 4, 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, 10A, RayPlan 1, 2, 7, 8A, 8B, 9A, 9B, 10A, including all service packs. If a region of interest (ROI) or point of interest (POI) that is referenced from an imported plan is missing in the imported RT Structure Set, the reference may become linked to the wrong ROI or POI. The imported RT Structure Set must be corrupt for the bug to be triggered, i.e., the referenced ROI/POI must not exist in the imported RT Structure Set. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/10/2020 |