Detail of Recall

Type of Product:Device
Medsafe Reference:26692
Brand Name:RaySearch_Radiation therapy treatment planning system
Model:
Affected:RayStation 4, 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, 10A, RayPlan 1, 2, 7, 8A, 8B, 9A, 9B, 10A, including all service packs
Software version:
Recalling Organisation:CARSL Consulting
30 Fulford Road
Havelock North 4130
Hastings
Contact Information:09 238 3447
Manufacturer:RaySearch Laboratories AB (publ)
Issue:The manufacturer has identified an issue found with the DICOM import in RayStation 4, 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, 10A, RayPlan 1, 2, 7, 8A, 8B, 9A, 9B, 10A, including all service packs. If a region of interest (ROI) or point of interest (POI) that is referenced from an imported plan is missing in the imported RT Structure Set, the reference may become linked to the wrong ROI or POI. The imported RT Structure Set must be corrupt for the bug to be triggered, i.e., the referenced ROI/POI must not exist in the imported RT Structure Set.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/10/2020