Detail of Recall

Type of Product:Device
Medsafe Reference:26669
Brand Name:Infinity Acute Care System Monitoring - M540
Model:
Affected:Software version VG7.1
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draegerwerk AG & Co KGaA
Brand Name:INFINITY Patient Monitors & Accessories
Model:
Affected:Software version VG7.1
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draeger Medical Systems Inc
Issue:Through post market activities the manufacturer has observed issues on the Drager Infinity Acute Care System (IACS) with software version VG7.1; and Infinity M540 Stand Alone Patient Monitor with software version VG7.1.
1. The Infinity M540 patient monitor may randomly reboot due to an error to correctly transmit and read the header data of files in the memory of the device.
2. The Infinity Acute Care System Cockpit may lose the ability to send commands to the Infinity M540 Patient Monitor.
3. The Infinity Acute Care System Patient Monitor may exhibit a loss of configuration of the patient and system profile due to disconnection from the internal database after power up of the Cockpit. If this problem is present, it will impact patient category, patient name, patient bed name, network settings, trends, and the units of measure.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/10/2020