Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 26652 |
| Brand Name: | Patient monitor, multiparameter |
| Model: | |
| Affected: | 13700259, 13700261, 13700693, 13700697, 13700864, 13701527, 13701528, 13701529, 13701892, 13702706, 13702719, 13710617, 13710621, 13710622 |
| Software version: | |
| Recalling Organisation: | Edwards Lifesciences (New Zealand) Limited 171 Morrin Road St John Auckland 1072 |
| Contact Information: | 021 313 234 |
| Manufacturer: | Edwards Lifesciences LLC |
| Issue: | The manufacturer has identified an issue that involves HemoSphere oximetry cables that have operated at a temperature which may cause an Oximetry Cable malfunction fault/error message and has the potential to stop producing Oximetry values. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/10/2020 |