Detail of Recall

Type of Product:Device
Medsafe Reference:26615
Brand Name:Gore TAG Thoracic Endoprosthesis (including Conformable)
Model:
Affected:All lot numbers
Software version:
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:W L Gore & Associates Inc
Issue:The manufacturer has received reports for two separate events for GORE TAG Conformable Thoracic Stent Graft with ACTIVE CONTROL System. The first was where the device remained at intermediate diameter after successful primary deployment. The secondary deployment line when removed did not actuate deployment of the device to full diameter and was not accessible via the Deployment Line Access Hatch. The second report where difficulty withdrawing the delivery catheter was encountered due to a proximal stent apex being lodged within the delivery catheter leading olive. Based on these events, the manufacturer will be updating its Instructions for Use (IFU) to include a new Warning
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/9/2020