Detail of Recall

Type of Product:Device
Medsafe Reference:26613
Brand Name:THE INSIDES System (fi)
Model:PS007
Affected:LOT0000006 and LOT0000007, LOT 5
Software version:
Recalling Organisation:The Insides Company Limited
VCFO Group Ltd
Level 1, 96 St Georges Bay Road
Parnell
Auckland 1052
Contact Information:
Manufacturer:The Insides Company Limited
Issue:The malecot wing shape design of The Insides Tube could create points of pressure which could cause granulation. This granulation may be able to grow into the wings of the malecot which could lodge the tube into the intestine.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/9/2020