Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 26613 |
| Brand Name: | THE INSIDES System (fi) |
| Model: | PS007 |
| Affected: | LOT0000006 and LOT0000007, LOT 5 |
| Software version: | |
| Recalling Organisation: | The Insides Company Limited VCFO Group Ltd Level 1, 96 St Georges Bay Road Parnell Auckland 1052 |
| Contact Information: | |
| Manufacturer: | The Insides Company Limited |
| Issue: | The malecot wing shape design of The Insides Tube could create points of pressure which could cause granulation. This granulation may be able to grow into the wings of the malecot which could lodge the tube into the intestine. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 23/9/2020 |