Detail of Recall

Type of Product:Device
Medsafe Reference:26609
Brand Name:Bien Air - Remote control, foot switch
Model:
Affected:Serial Numbers: 17F0060, 18F0022, 17F0013
Software version:
Recalling Organisation:Downs Distributors Ltd
9 Airborne Road
North Harbour
AUCKLAND
Contact Information:0800 800 234
Manufacturer:Bien Air Dental SA
Issue:The manufacturer advises that there is a low possibility that the magnet located inside the foot pedal displaces or comes off during a surgery and therefore may compromise the device's safety. Use of an affected device may entail a risk of unattended start or unstoppable motor when acting on the foot pedal.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/9/2020