Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 26572 |
| Brand Name: | Mobile OR Table |
| Model: | MEERA US w/o Drive Unit - 221 to 226 and 701 to 711, Mobile OR-Table YUNO OTN EU - 1 to 267, Mobile OR-Table YUNO II EU - 1 to 62 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Getinge Australia Pty Ltd (NZ) Unit D, Ground Floor, Building A, Millenium Centre 600 Great South Rd Ellerslie Auckland 1051 |
| Contact Information: | 09 272 9039 |
| Manufacturer: | Maquet GmbH |
| Issue: | If the mains cable (power cord) is still connected to the mains socket (power source), there is a risk of electric shock to user or third parties when touching exposed, electronic components. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 16/9/2020 |