Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:26560
Brand Name:Coated VICRYL (Polyglactin 910) Suture
Model:
Affected:VICRYL MONOFILAMENT VIOLET 4IN (10CM) USP10-0 (M0.2) SINGLE-ARMED CS140-6 - LOT - PJ5352
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Ethicon Inc
Issue:The manufacturer has identified a foil package labelled VICRYL V960, Lot PJ5352 that may contain VICRYL J551, Lot PJ5055. The correct VICRYL suture is size 10-0, 4 inches (10 cm) in length, has monofilament construction, and a single needle. The incorrect VICRYL suture is size 6-0, 18 inches (45 cm) in length, has braided construction, and two needles.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/9/2020
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /