Detail of Recall

Type of Product:Device
Medsafe Reference:26523
Brand Name:MITG - Pipeline Shield Embolization Device
Model:
Affected:Pipeline™ Flex Embolization Device and Pipeline™ Flex Embolization Device with Shield Technology
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Micro Therapeutics Inc dba ev3 Neurovascular
Issue:The manufacturer is providing a labelling update to ensure awareness among Pipeline™ Flex and Pipeline™ Flex with Shield Technology™ users, which describes the existing risks and potential patient harms associated with separation or fracture. This labelling change further clarifies certain conditions where the risk of separation may be increased, such as under conditions of increased vessel tortuosity and/or excessive resistance during advancement or retraction of the device.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/9/2020