Detail of Recall

Type of Product:Device
Medsafe Reference:26448
Brand Name:ArthroCare - Suture/Needle Passer
Model:
Affected:Lot no :2018434, 2018934, 2019558, 2019559, 2019929, 2019930, 2020931, 2022111, 2022641, 2030008, 2032212, 2032526, 2033815, 2036941, 2037604, 2038701, 2039699, 2040500, 2042295, 2044957, 2047483, 2048575 & 2051239
Software version:
Recalling Organisation:Smith & Nephew Surgical Ltd
Unit A 36 Hillside road
Wairau Valley
AUCKLAND 0760
Contact Information:0800 657 799
Manufacturer:ArthroCare Corporation
Issue:Due to a supplier error where a portion of the raw material was inadvertently mixed, as a result, some needles may be made of commercially pure titanium as opposed to 304 stainless steel which does not meet the specification and could potentially break during use.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/8/2020