Type of Product:Device
Medsafe Reference:26378
Brand Name:Alinity m System
Model:
Affected:All lot numbers
Software version:
Recalling Organisation:Abbott Laboratories (NZ) Ltd
Unit D
4 Pacific Rise
Mt Wellington
Auckland 2025
Contact Information:0800 656 233
Manufacturer:Abbott Laboratories (NZ) Ltd
Issue:There is potential impact to sample processing when using gel tubes for Alinity m HIV-1, HCV and HBV testing. If a gel tube is dropped inside the instrument, there is a potential for a delay of results. Although improbable, there is potential for biohazard exposure or cross-contamination.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/8/2020