Detail of Recall

Type of Product:Device
Medsafe Reference:26364
Brand Name:Elecsys Anti-TPO
Model:
Affected:All lot numbers
Software version:
Recalling Organisation:Abbott Laboratories (NZ) Ltd
Unit D
4 Pacific Rise
Mt Wellington
Auckland 2025
Contact Information:0800 656 233
Manufacturer:Abbott Diagnostics
Issue:The issue as described leads to a maximum recovery of 125.9 % for Anti-TPO at 29.5 IU/mL. The cutoff value for Anti-TPO is given at 34 IU/mL. Falsely elevated Anti-TPO results may lead to wrongly assumption of positive Anti-TPO status of a patient and to wrong assumption of Hashimoto's thyroiditis in patients as sublying cause of (subclinical) hypothyroidism and may affect further diagnostic steps.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/8/2020