Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 26359 |
| Brand Name: | ACUSON SC2000 |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Medical Solutions USA Inc |
| Issue: | The eSie Flow analysis algorithm is not able to produce a reliable and accurate result due to variations in color Doppler data being used as input. The decision has been made to discontinue commercial distribution of the eSie Flow quantification package. This also leads to the decision to remove the feature license from existing customer's systems. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 31/7/2020 |