Detail of Recall

Type of Product:Device
Medsafe Reference:26355
Brand Name:External Defibrillator (AED)
Model:
Affected:M3535A, M3536A
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The manufacturer has identified that the therapy switch may fail, resulting in the device exhibiting the following behaviors:
The device may not perform the selected function.
The therapy knob may not change to the energy setting selected.
The device may deliver a shock with an energy level different from the setting selected by the user.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/7/2020