Detail of Recall

Type of Product:Device
Medsafe Reference:26346
Brand Name:Troponin I (TNIH) Flex® reagent cartridge
Model:
Affected:Lot numbers - 20008BB, 20035BC, 20135BB
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has observed a negative bias across the Analytical Measurement Range of the Dimension Vista High Sensitivity Troponin I (TNIH) assay. The average bias observed for patient samples using the Dimension Vista TNIH lots when compared with a control TNIH lot was -20% for lot 20008BB and -23% for lot 20135BB. A maximum bias of -26% in patient samples around the 99th percentile with the lots 20008BB, 20035BC and 20135BB was observed. Quality Control will not detect the negative patient bias. TNIH lot 20008BB and lot 20035BC contain the same reagents therefore, performance of the two lots is equivalent.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/7/2020