Detail of Recall

Type of Product:Device
Medsafe Reference:26321
Brand Name:Physio-Control Inc - Rechargeable public semi-automated external defibrillator
Model:for all LIFEPAK 500 AEDs
Affected:
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Physio-Control Corporation
Issue:The manufacturer has become aware that LIFEPAK 500 AEDs in high-use environments (Emergency Medical Services) may not detect a patient connection due to mechanical wear-through of the contact plating on the therapy connector. Wear-through of the connector exposes base metal on which an oxide layer may form and result in the device not recognising a patient is connected. When this circumstance arises, the device will provide the user the "CONNECT ELECTRODES" message.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/7/2020