| Medsafe Reference: | 26309 |
| Brand Name: | EVO IQ LVP |
| Model: | |
| Affected: | Product Code ELVP001ANZ - All versions V01.02.00.04 |
| Software version: | |
| Recalling Organisation: | Baxter Healthcare Ltd 33 Vestey Drive Mt Wellington AUCKLAND |
| Contact Information: | 0800 229 837 |
| Manufacturer: | Baxter Healthcare SA (Switzerland) |
| Issue: | The manufacturer has identified an issue with the EVO IQ Large Volumetric Pump (LVP) due to a software error which affects the optional drug modifier feature. The error occurs when multiple drugs within a care area start with the same first letter and have the same exact drug modifier names. If these conditions exist, when one of these drugs are selected on the pump, the available options displayed within that modifier (concentrations, dose modes, etc.) are for multiple drugs, and it is not clear to the user that some of the selections available do not apply to the drug that has been selected. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/7/2020 |