Type of Product:Device
Medsafe Reference:26307
Brand Name:Rocket Fetal Blood Sampling Kits
Model:
Affected:Product Code: R57028-00-SW - All LOT numbers inclusive from 000000000475490 to 000000000489970
Software version:
Recalling Organisation:Rocket Medical (NZ) Ltd
Sidekick Auckland North Ltd
Level 1
445 Lake Road
Takapuna 0622
Contact Information:+64 021 198 8488
Manufacturer:Rocket Medical PLC
Issue:The device is indicated for use in lactate analysis of fetal blood, the manufacturer has become aware that the tubes within the kit are not of the required specification for lactate analysis using some analysers, and cannot rule out that this may result in an incorrect / misleading reading from which a clinical decision is made.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/7/2020