Type of Product:Device
Medsafe Reference:26304
Brand Name:BD MAX™ SARS-CoV-2 Reagent 445003
Model:
Affected:All Lots
Software version:
Recalling Organisation:Becton Dickinson Pty Ltd
80 Rushdale Street
Knoxfield
VIC 3180
Australia
Contact Information:
Manufacturer:Becton Dickinson and Company (USA)
Issue:The manufacturer has received reports of potential false positive results when using the BD SARS-CoV-2 Reagents for the BD MAX™ System. A false positive SARS-CoV-2 test result could lead to incorrect triaging for that patient. On review, the manufacturer is revising the User Defined Protocol (UDP) threshold for the N2 target in the Instructions for Use (IFU) from 50 RFU to 80 RFU.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/7/2020