Detail of Recall

Type of Product:Device
Medsafe Reference:26298
Brand Name:Radiology picture archiving and communication system application software
Model:
Affected:SICAT IMPLANT V2.0 software
Software version:
Recalling Organisation:Dentsply Sirona (NZ) Limited
c/o Lowndes Jordan
Level 15, HSBC Tower
188 Quay Street
Auckland 1010
Contact Information:
Manufacturer:SICAT GmbH & Co KG
Issue:In certain cases, the error leads to the planned implant-sleeve combination being exported shifted by a positive or negative offset along the implant axis. The offset is the difference between the nominal implant length1 and the realistic implant length2. The sign of the offset also depends on the format of the import data (SSI, SIXD). The exported CMG.DXD file is incorrect in these cases.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Healthcare Professional
Date Commenced:17/7/2020