Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:26286
Brand Name:Cios Alpha / Fusion
Model:
Affected:Material number(s): 10308194 & 10308191
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:The manufacturer has identified that hot plugging feature could cause an electrical malfunction on the main control D80 of the generator which is part of the main unit (c-arm). If malfunction occures during hot plugging, this will cause a permanent loss of imaging functionality and interruption of the clinical procedure.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/7/2020