Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:26193
Brand Name:EliA ANA Positive Control 250
Model:
Affected:PU83103301 MR13/C9RDF is the only lot affected in NZ
Software version:
Recalling Organisation:Abacus ALS
12 Mowbray Terrace
East Brisbane
QLD 4169
Australia
Contact Information:
Manufacturer:Phadia AB
Issue:The investigation shows that many of the results are found in the lower half of the expected range for EliA dsDNA Well of the affected lots. Consequently, more results may be found below the limit stated in the certificates. This does not have an effect on the correctness of the test results generated for patient samples since the EliA dsDNA wells are not affected.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/6/2020