Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 26167 |
| Brand Name: | Patient monitor module, temperature |
| Model: | CARESCAPE B450/B650/B850 host monitor |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems Information Technologies Inc |
| Issue: | If a patient presents with a V Fib arrhythmia at the time CARESCAPE ONE is docked to a CARESCAPE B450/B650/B850 host monitor, and the host monitor is in the Learning process from CARESCAPE ONE, the host monitor may not alarm for a V Fib arrhythmia. In rare circumstances, this issue could result in delayed clinical assessment of a V Fib arrhythmia. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 23/6/2020 |