Detail of Recall

Type of Product:Device
Medsafe Reference:26167
Brand Name:Patient monitor module, temperature
Model:CARESCAPE B450/B650/B850 host monitor
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems Information Technologies Inc
Issue:If a patient presents with a V Fib arrhythmia at the time CARESCAPE ONE is docked to a CARESCAPE B450/B650/B850 host monitor, and the host monitor is in the Learning process from CARESCAPE ONE, the host monitor may not alarm for a V Fib arrhythmia. In rare circumstances, this issue could result in delayed clinical assessment of a V Fib arrhythmia.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/6/2020