Detail of Recall

Type of Product:Device
Medsafe Reference:26122
Brand Name:KCI (IH10) -V.A.C. Therapy Unit (including InfoVac & Activac)
Model:340003 ACTIV.A.C.™ Therapy Unit (All serial numbers)
Affected:
Software version:
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:KCI USA Inc
Issue:The manufacturer has become aware that ACTIV.A.C™ Therapy Units may power off without notification to the user (i.e., no alarm or warning) resulting in a stoppage of negative pressure wound therapy.Ssince April 2017, the manufacturer has received reports of injury, which have resulted in maceration, localised infection, or wound deterioration as a result of the ACTIV.A.C™ Therapy Unit unintentionally powering off.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/8/2022