Detail of Recall

Type of Product:Device
Medsafe Reference:26109
Brand Name:InteleradPACS
Model:
Affected:InteleViewer versions 4-10-1 and up
Software version:
Recalling Organisation:Pharmaceutical Marketing New Zealand Ltd
162 Nile Road
Milford
Auckland 0620
Contact Information:09 377 0660
Manufacturer:Intelerad Medical Systems Incorporated
Issue:The Automatic Image Registration tool may be failing image registration when it is used with non-DICOM compliant JPEG 2000 compressed images. The software improperly uses the DICOM bit depth (i.e. the Bits Stored tag) to decompress the compressed pixel data stream for analysis, instead of the bit depth that is encoded in the compressed pixel data stream itself
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/6/2020