Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 26098 |
| Brand Name: | Hemochron Signature Elite_Accriva |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Getz Healthcare Limited Level 17 120 Albert Street Auckland |
| Contact Information: | 0800 733 333 |
| Manufacturer: | Accriva Diagnostics |
| Issue: | The manufacturer has identified that some Hemochron Signature Elite analysers may not maintain a temperature of 37°C ±1.0°C degrees in some operating environments as outlined in the Operators Manual (P/N HX1101). |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 10/6/2020 |