Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:26098
Brand Name:Hemochron Signature Elite_Accriva
Model:
Affected:
Software version:
Recalling Organisation:Getz Healthcare Limited
Level 17
120 Albert Street
Auckland
Contact Information:0800 733 333
Manufacturer:Accriva Diagnostics
Issue:The manufacturer has identified that some Hemochron Signature Elite analysers may not maintain a temperature of 37°C ±1.0°C degrees in some operating environments as outlined in the Operators Manual (P/N HX1101).
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/6/2020
Hide menus
Show menus