Detail of Recall

Type of Product:Device
Medsafe Reference:26090
Brand Name:Arrow International - Intra-aortic Balloon Pump
Model:
Affected:Serial number - 190336F
Software version:
Recalling Organisation:Teleflex Medical New Zealand
OFFICE 011, 110 CARLTON GORE ROAD
Newmarket
Auckland 1023
Contact Information:027 476 4009
Manufacturer:Arrow International Inc
Issue:A potential issue with a component ("Component") within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for "System Error 3" and "High Baseline" presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP. The potential for this issue to occur is related to aging of the Component
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/6/2020