Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 26088 |
| Brand Name: | Cryotherapy Needles |
| Model: | IceForce 2.1 CX and IcePearl 2.1 CX Cryoablation Needles |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Urotech Limited 706 Wright Road RD 4 Katikati Tauranga 3181 |
| Contact Information: | |
| Manufacturer: | Galil Medical |
| Issue: | Following one report of a serious injury in a patient during a procedure due to a high-pressure gas leak from a needle during the electrical thawing operation. The initial investigation has found that the needle surface deteriorated due to unexpected electrolysis during the procedure. The duration of the electrolysis was long enough to cause a perforation in the shaft wall. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/6/2020 |